OPERATING ENVIRONMENT
Ambient temperature 15-30oC
Atmospheric pressure 860-1060 hPa
Relative humidity 30-75% (non-condensing)
Measurement range 2-200 ppb
Sampling frequency 20 Hz
Lag time <0.8 sec
Rise time 10-90% <0.7 sec
Response time <1.5 sec
Fall time 90%-10% <1 sec
Offset ± 3 ppb/14 days
Accuracy ± 2.5 ppb of measured value <50 ppb ± 5 % of measured value >50 ppb
Precision <2.5 ppb of measured value <50 ppb (analytical and clinical precision)
<5% of measured value >50 ppb (analytical precision)
(expressed as standard deviation for concentrations below 50 ppb or as a coefficient of variation of measured value for concentrations above 50 ppb)
Linearity
<2.5 ppb integral linearity
Detection Level 2 ppb
Interference <10 % of measurement value at 5% volume H2O and 5% volume CO2
Calibration interval Recommended every 14 days, or whenever system is restarted or moved. Recommended calibration gas NO in N2: 200 ± 20 ppb 3 bar absolute.
NIOX® Filter Disposable filters to be changed for every new patient. Bacterial and viral filter according to Medical Device Directive Class IIa.
Scheduled
service
NO-scrubber to be replaced every 6 months. Scheduled service of the entire NIOX system should be performed every 18 months.
Dimensions and weight Height 500 mm
Width 300 mm
Depth 400 mm
Weight 40 kg without gas cylinder for calibration
Electrical data Certified by Intertek-SEMKO according to EN/IEC 60601-1
ETL-Classified according to General Requirements for Safety CAN/CSA-C22.2 No. 601.1 and UL-60601-1
Class I Type B
EMC approved class B
Consumption: 2 A (at 220 V)
Mains voltage: 100–240 V ~ 50–60 Hz
Fuse: T5 AL 250 V
Software Software platform: Microsoft® Windows XP® Professional
Application Software: NIOX® Software